Microscopy
Introduction
Microscopes are useful for viewing objects that are too small to
see clearly without magnification. This exercise is designed to familiarize
students with the use of a compound light microscope and a binocular dissecting
microscope.
Before using the microscope, read the document on microscope
care using the link below.
http://faculty.clintoncc.suny.edu/faculty/Mike.Lawliss/LabRefandOrient/Microsccope.htm
Compound Light Microscope
The compound light microscope uses two sets of lenses to magnify the object.
Illumination is provided by a light source on the base of the microscope. The
magnification typically ranges from approximately 40 X to 1,000 X. They can be used with
objects that range in size from about 100 nm to 2 mm.
Parts of the Light Microscope
The stage is a platform that holds the slide containing
the specimen to be viewed. A mechanical stage (see the photographs below)
has a mechanism for moving the slide.
A light microscope must have a light source. This is usually a
light bulb located beneath the stage.
An adjustable diaphragm located beneath the stage is used to
regulate the amount of light that passes through.
A condenser contains two sets of lenses that concentrate light.
It is located directly underneath the stage. Light from the light source passes
through the diaphragm and condenser before continuing up through the specimen to
be viewed.
The body tube contains an ocular lens
(eyepiece) and a nosepiece with several objective lenses.
Each objective lens is used for a different magnification and is moved into
place by rotating the nosepiece. The image is brought into focus by adjusting the coarse and fine focus
knobs.
 |
Click on the photographs to view enlargements.
After viewing, press the "Back" key to
return here.
|

Binocular microscopes (below) have two eyepieces; monocular microscopes
(above) have
one. The distance between the two ocular lenses of a binocular microscope can be adjusted to fit the
distance between your eyes.
Other optical devices such as binocular telescopes and field glasses also have two ocular lenses that
adjust in a manner similar to the microscope. Binocular lenses can be adjusted
individually, making it unnecessary for many people to need their glasses when
using them. If you wear glasses and are unfamiliar with adjusting binocular lenses
to correct for your own eyes,
see the section titled "Adjusting the Ocular Lenses"
below.
|
Click on the photographs to view enlargements. |
Theory
Compound light microscopes contain two lens systems, an objective and an ocular. The total magnification of an image
is calculated by multiplying the magnification of the ocular by the
magnification of the objective. The microscopes we will use each have a 10X ocular
lens and four different objective lenses listed in the table below.
| Objective |
Magnification |
Total Magnification |
| Scanning |
4 X |
40 X |
| Low Power |
10 X |
100 X |
| High Power |
40 X or 43 X |
400 X or 430 X |
| Oil Immersion |
100 X |
1000 X |
Light bends when it passes from glass to air or from air to glass because
air and glass have different refractive indices. The bending of light as it
passes through the glass slide to the air and then to the glass lens decreases
the resolving power. At high magnification (1000X) it can prevent a clear
image from being viewed. This decrease in resolution can be prevented by putting immersion
oil between the slide and the lens because immersion has the same
refractive index as glass.
The condenser also increases the resolving power of the microscope. When
using the oil-immersion lens, the condenser (located beneath the stage) should
be raised to a position very close to the stage for maximum resolution.
The link below may be helpful before using the microscope.
http://www.udel.edu/biology/ketcham/microscope/
CAUTION - Never use cloth or paper products (paper towels,
tissue paper, etc.) to clean the lenses. They will scratch the coating and
decrease the resolving power of the lens. Use only lens
paper.
Switch the microscope to the lowest magnification or raise the objectives
from the stage before inserting a slide. This will prevent the objective lens
from being accidentally scratched by the slide.
Place the slide to be viewed on the stage and center the specimen over the
opening.

Begin with either the scanning lens or the low power objective lens.

Raise the stage (or lower the lens) all the way so that the slide is as
close as possible to the objective lens.
Use the coarse adjustment know to slowly raise the lens from the stage
while viewing the image. Fine focusing is not needed when using the lowest
magnification (scanning or 4X objective). If you are using any of the other
objectives, it will be necessary to use the fine focus after using the coarse
focus.
Adjust the condenser so that a sharp focus is produced. This step is
important at the highest magnification (oil immersion or 1000X).
Adjust the iris diaphragm. This will need readjustment after changing to a
different magnification.

To Increase the Magnification
The microscopes are parfocal, meaning that after you adjust
the focus, the image will remain approximately in focus if you change the
magnification.
Center the object before switching to a higher power objective. This will
help you find the object after switching the objective.
Switch to the next highest power. It will be necessary to center the image
again. The image should be approximately in focus but it will be necessary to use
the fine focus. The coarse focus should not be needed after switching
objectives.

Adjust the diaphragm.
This procedure is repeated each time you switch to a higher magnification.
The 100 X objective (1,000X total magnification) requires that a drop of
immersion oil be placed between the slide and the lens. We will not use
this objective but students are welcome to try the lens if they would like.
The procedure is given below.
After focusing the specimen under high power (400X or 430X, see above), rotate
the high power objective out of the way and place a drop of immersion oil on
the slide. Rotate the oil immersion objective into place so that it touches
the oil.
Adjust the fine focus, condenser, and iris diaphragm as previously
described.
After viewing with oil, the lens must be cleaned with fluid designated for
this purpose. Remember, use lens paper only. Never use cloth, paper
towels, or other paper products on coated optics.
Practice Using the Microscope
- Obtain a slide of colored threads and view them under the scanning and low
power. Use the focusing procedure described above.
1) Can you tell which thread is above the other?
View the threads under high power (400X or 430X). Use the fine focus to focus to
determine the order of the threads from top to bottom. As you rotate the fine
focus, different strands will go out of focus while others will become more
sharply focused. This procedure will therefore enable you to determine the order
of the threads.
2) Are all of the threads in focus at the same time?
3) What is the order (from top to bottom)?
"Depth of field" refers to the thickness of the plane of focus.
With a large depth of field, all of the threads can be in focused at the same
time. With a smaller or narrower depth of field, only one thread or a part of
one thread can be focused, everything else will be out of focus. In order to
view the other threads, you must focus downward to view the ones underneath and upward to view the
ones that are above.
4a) What happens to the depth of
field when you increase to a higher magnification (increases, decreases, or
remains the same)?
4b) Explain how the slide with threads could be used to answer the question
above.
|
Click on the image to view an enlargement. |
- Obtain a slide of the letter e and view it under the scanning objective. Move
the slide to the left.
5) What way did the image move when the slide was moved to the left?
6) How does the orientation of the image compare to the image on the slide?

Paramecium
- Obtain a prepared slide of paramecium and view it using the scanning lens
(40X). After the light and focus are adjusted, center a paramecium and
increase the magnification to 100X. Next, adjust the lighting, use the fine
focus, and then center the paramecium. Increase the magnification to 400X
and then adjust the light and adjust the fine focus.
- The procedure described above (adjust the light, adjust the focus, center
the specimen, increase the magnification) should be used whenever you are
trying to view a small specimen that is difficult to find.
7) Draw the cell. Label the cilia and the oral groove.
Below: Paramecium.
 |
Click on the image to view an enlargement. |
Making Wet Mounts
Wet mounts are useful for viewing living biological
material.
To make a wet mount, place the specimen on a slide
and then add a drop of water or stain. Stain is often used to make the
specimen more visible.
|
Click the photographs to view enlargements. |
Place a cover slip at an angle so that it touches
the drop. Slowly lower the raised end of the cover slip. The diagram below
shows that as the cover slip is lowered, the drop of liquid moves to the
right.


Cheek Cells
- Scrape the inside of your cheek with a toothpick and rub it on a dry
slide.
- Add one drop of methylene blue to stain the cells. This will make them
easier to see.
- Place a cover slip on the slide as described above and observe the cells
under low power then high power.
1) Draw a cell under high power.
Below: Cheek cells 100X and 400X.
|
 |
Click the photographs to view enlargements. |
Onion Cells
- Prepare a wet mount of fresh onion and stain it with methylene blue. Try
to get the thinnest piece that you can; it should be thinner than paper. It
might help if you take a thicker piece and bend it until it snaps into two
pieces. After it snaps, there is usually a very thin connection remaining
between the two pieces. This thin connecting layer can be peeled off and
placed on a microscope slide for viewing. After you view this slide and answer the questions below, set the slide
aside for use later in the laboratory period.
2) Draw several onion cells as they appear under high power.
Below: Onion cells.
|
Click on the image to view an enlargement. |
Binocular microscopes and stereomicroscopes have one ocular lens that is
adjustable (see photograph below). This enables you to adjust the viewing for
your eyes so that you do not need to wear your glasses. Remove your glasses
before doing the procedure below.
To adjust these lenses, first, cover the adjustable lens (or your eye) so
that you cannot see the image through it. Next, focus the microscope the way you
normally would so that a sharp image is produced through the ocular lens. Cover
this lens (or eye) and view the image through the other (adjustable) ocular. Turn the
ocular lens but not the focus knob and adjust so that the image is sharp.

|